Quantification of the uranium concentration in human urine by inductively coupled plasma-sector field mass spectrometry (ICP-SFMS) Article uri icon

abstract

  • In this study, the validation of a method for analyzing the uranium (U) concentration in human urine samples by inductively coupled plasma-sector field mass spectrometry (ICP-SFMS) was conducted. PROCORAD (the Association for the Promotion of Quality Control in Radiotoxicological Analysis) provided two urine samples spiked with unknown contents of U (Sample A = 33.6 ± 1.0 μg/L and Sample B = 3.3 ± 0.1 μg/L) and one unspiked sample as a blank. The analyses were directly performed on the diluted urine samples (dilution factor = 1:20) in 5%25 v/v HNO3. The results obtained by ICP-SFMS corresponded well with the reference values, and the limits of detection were 235U = 0.049 × 10-3 μg/L and 238U = 7.37 × 10-3 μg/L. The ICP-SFMS technique has been shown to be successful in the analysis of the U concentration in human urine samples and for the quantification of isotopic ratios.

publication date

  • 2013-01-01