UPLC–MS/MS method for fluconazole determination in plasma and its application in Mexican patients with candidaemia Article uri icon

abstract

  • Aim: An accurate and fast ultra-high performance liquid chromatography coupled with tandem mass spectrometry analytical method was developed and validated for quantifying fluconazole levels in human plasma according to the US FDA guidelines. Materials & methods: A simple protein precipitation by acetonitrile was employed for the sample preparation. The chromatographic separation was carried out using isocratic elution of water (0.1%25 formic acid) and acetonitrile on an Acquity ultra-high performance liquid chromatography HSS T3 column. Samples from ten adult patients diagnosed with candidemia who received fluconazole treatment were analyzed. Results & conclusion: The method demonstrated excellent linearity and stability within the 1–50 μg/ml range (r2 >0.999). The intraday and interday precision were determined with coefficient of variation values ranging from 1.4 to 4.38%25 and 2.8 to 6.6%25, respectively. This rapid and selective method has successfully analyzed 27 plasma samples. The straightforward sample preparation in a single step and the reduced analysis time make this method suitable for adult patients with candidemia, leading to improved clinical outcomes.

publication date

  • 2024-01-01